  • Programmable humidity and temperature transmitters with 4-20mA output

    photo of humidity transmitters

    • temperature
    • relative humidity
    • dew point temperature
    • absolute humidity
    • specific humidity
    • mixing ratio
    • specific enthalpy

    Programmable temperature and humidity transmitters are equipped with temperature and relative humidity sensors. Measured values are also converted to other humidity interpretation - dew point temperature, absolute humidity, specific humidity, mixing ratio or specific enthalpy. Degrees Celsius and Fahrenheit are user selectable. Transmitters are available in wall-mount, duct-mount and bar types. Also types with T+RH probe on a cable are available. Transmitter contains a microprocessor based control circuitry in a durable plastic case with connection terminals and sensors in a filter with stainless steel mesh. Humidity transmitters are also available with two galvanic isolated 4-20mA outputs. Configuration of outputs and output range are user adjustable. Large dual line LCD for simultaneous display of T+RH or other humidity interpretation is an advantage. Display is possible to switch off. Computerized design ensures temperature compensation of the humidity sensor and fail indication. State-of-the-art capacitive polymer sensor ensures excellent calibration long term stability, inertia against water and condensation. Transmitters are designed for use in non-aggressive environment.

    APPLICATIONS - measuring and control of air humidity at:

    • warehouses
    • glasshouses
    • manufacturers
    • museums, archives, galleries
    • air-conditioned rooms
    • weather stations
    COMMON TECHNICAL PARAMETERS
    Air relative humidity range: 0 to 100%
    Accuracy of relative humidity measurement: ±2.5% relative humidity from 5 to 95% at 23°C
    Accuracy of temperature output: ±0.4°C from -30 to +100°C
    ±0.4% from reading over +100°C
    Available temperature units: degress Celsius, Fahrenheit
    Accuracy and range of dew point temperature output: ±1,5°C at ambient temperature T<25°C and RH>30%
    range -60 to +80 °C
    Accuracy and range of absolute humidity output: ±3g/m3 at ambient temperature T < 40°C
    range 0 to 400 g/m3
    Accuracy and range of specific humidity output: ±2g/kg at ambient temperature T < 35°C
    range 0 to 550 g/kg
    Accuracy and range of mixing ratio output: ±2g/kg at ambient temperature T < 35°C
    range 0 to 995 g/kg
    Accuracy and range of specific enthalpy output: ±3kJ/kg at ambient temperature T < 25°C
    range: 0 to 995 kJ/kg
    Temperature operating range of case with electronics: -30 to +80°C
    Temperature operating range of LCD display: readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C
    Range of humidity sensor temperature compensation: -30 to +125°C
    Current outputs - two-wire connection: 4-20mA, galvanic isolated with dual-output models
    Configuration of outputs and output range: user adjustable from the PC
    Filtering ability of sensor cover: 0.025mm
    Protection of the case with electronics: IP65 electronics, IP40 sensors
    Power: 9-30Vdc
    Warranty: two years

    Humidity transmitters are available in the following types:

    TypeMeasured valueMaximum range of temperature measurementLengthOutput 1**Output 2**Note

    T1110

    Larger photo of humidity transmitter
    RH -30 to +80°C 75mm 0 to 100%RH - Outdoor, indoor use.
    Only relative humidity output.

    Mechanical drawing

    Interconnection terminals

    Manual in pdf
    Humidity transmitter - manual in pdf

    T3110

    Larger photo of humidity transmitter
    RH+T -30 to +80°C 75mm 0 to 100%RH -30 to +80°C outdoor, indoor

    Mechanical drawing

    Interconnection terminals

    Manual in pdf
    Humidity transmitter - manual in pdf

    T3111

    Larger photo of humidity transmitter
    RH+T -30 to +105°C* probe including cable probe cable 1,2,4m 0 to 100%RH -30 to +105°C T+RH probe with 1m cable. Cable lengths 2m or 4m available.

    Mechanical drawing

    Interconnection terminals

    Manual in pdf
    Humidity transmitter - manual in pdf

    T3113

    Larger photo of humidity transmitter
    RH+T -30 to +125°C* 150mm 0 to 100%RH -30 to +125°C duct mount

    Mechanical drawing

    Interconnection terminals

    Manual in pdf
    Humidity transmitter - manual in pdf

    T3117

    Larger photo of humidity transmitter
    RH+T -30 to +125°C* 700mm 0 to 100%RH -30 to +125°C bar type

    Mechanical drawing

    Interconnection terminals

    Manual in pdf
    Humidity transmitter - manual in pdf

    * Maximum temperature only at the measuring end with sensors. Maximum temperature +105°C for T3111 with cable probe is allowed also for the cable. Relative humidity at temperature over +85°C is limited in accordance with the graph. Near plastic case with electronics maximum temperature is +80°C.

    ** Any measured value - temperature, relative humidity, dew point temperature, absolute humidity, specific humidity, mixing ratio or specific enthalpy can be assigned to each output of dual output transmitters. Also identical value can be assigned to both outputs. Outputs are adjusted to maximum range from the manufacturer. Output range is user adjustable from the PC by means of the optional cable SP003 - see below. Free configuration program for transmitter adjustment is ready to download anytime. If different adjustment of outputs and output ranges are required, please specify required output values (RH, T, Tdp, ..) and required ranges.

    Ordering example: Transmitter T3110, output 1: RH 10 to 90%, output 2: temperature 0 to 35°C

    Working range of humidity probe

    Figure: Working range of humidity probe

    humidity transmitter wiring humidity transmitter wiring

    Typical application wiring of single output nad dual output transmitters

    Calculation of minimum power voltage Uss for proper operation:
    Uss min> Uo min + Imax * Rz,
    where: Uo min= 9 V
    Imax... approximately 20 mA
    Rz... sensing resistor (shunt)
    Both current loops (I1 and I2) are connected analogically, current loop I1 must be always connected.

    Included accessory:

    Calibration certificate from the manufacturer, instruction manual. Free configuration program for transmitter adjustment is ready to download anytime.

    OPTIONAL ACCESSORY
    Larger photo of humidity transmitter cable SP003 cable for humidity transmitter adjustment via USB port
    Larger photo of humidity transmitter cover F5200 grey sensor cover with filter from stainless steel mesh, filtering ability 0,025mm for transmitters T1110, T3110, T3113, T3117
    Larger photo of humidity transmitter cover F5200B black sensor cover with filter from stainless steel mesh, filtering ability 0,025mm for transmitter T3111
    Larger photo of humidity transmitter cover F0000 sintered bronze sensor cover. Filtering ability 0.025mm.
    Larger photo of humidity transmitter flange PP4 flat plastic circular flange for duct mounting
    Larger photo of humidity transmitter flange PP90 right-angled stain-less steel flange for wall mounting
    Larger photo of humidity transmitter gland SP004 plastic gland for direct mounting of the humidity probe to a 29 mm diameter hole
    Larger photo of humidity transmitter Dual Lock MD036 self adhesive Dual Lock for easy installation
    Tool for easy transmitter connection - larger photo SP005 Tool for easy wire connection to terminals
    Solar radation shield Solar radation shield
    humidity calibration accessory for calibration and adjustment of humidity transmitter by the user

    Humidity transmitter is directly compatible with sixteen channel Comet data acquisition system MS5D, MS5.